Adjusting the alignment of the transmission lamp
1 • If the self-test shows that the transmission lamp is not properly aligned,
insert the special plastic-covered screwdriver (supplied with the scanner)
in the space between the scanner lid and the lamp shield to move the
lamp assembly
very slightly
(ca. 1 mm) in the indicated direction:
front end to the right = apply screwdriver at position 1
rear end to the right = apply screwdriver at position 2
front end to the left = apply screwdriver at position 3
rear end to the left = apply screwdriver at position 4
CAUTION: Use the screwdriver as a lever. Do not twist it.
2 • After making the adjustment, click on “Continue” to repeat the
alignment test.
3 • If correct alignment cannot be achieved after several tries, please
contact your ScanView dealer for assistance.
Maintenance
ScanMate F10 • Operator’s Manual
6•5
1
2
3
4
If adjustment is necessary, use the flat side of
the screwdriver as shown to move the lamp
assembly very slightly in the required direction.
Do not twist the screwdriver.